DTVB-200 To DTVX-2000 Tension Meters(Electronic) Model: DTVB-200 To DTVX-2000
Frequency
12518
Model No:DTVB-200 To DTVX-2000
Product Name:Tension Meters(Electronic) Model: DTVB-200 To DTVX-2000
Maker:distributor
Distributor:ABTEK CORP.
Price:contact us
Specification
4 Tension ranges from 2 - 200 cN to 200 - 2000 cN
Model for easy reading the measuring values, if filament run demand this
Tension meter with measuring head rotated by 90º
Tension meter with small measuring head, where access space is limited and turned-up outer finger edges for easy thread
Special Features
Turned-up outer finger edges guide the running filament into the roller grooves
Length of measuring head approx. 59 mm
Hand-held unit with small, ball-bearing mounted, V-grooved guide rollers
Tension meter model DTVX is calibrated using only PA monofilament!
Calibration to customer supplied material is available
The design is otherwise according to model DTMB respectively model DTMX, but no material thickness compensator
For additional Information, see model DTMB and DTMX
Standard Features
Everything in operator’s view: - the guide rollers - the measured material - the readings
Roller shifting mechanism ensure easy acquisition of the running material
Ball-bearing mounted, V-grooved guide rollers
Tension meter with rugged aluminium housing
Battery operated (AC adapter for continuous operation available)
Certificate of Compliance with the order 2.1 according EN 10204 is included
Optionally available: Inspection Certificate 3.1 according EN 10204 with calibration report
Available Models
Model
Model
Tension Range cN
Resolution
Measuring Head Width* mm
SCHMIDT Calibration Material PA**
DTVB-200
DTVX-200
2.0 - 200.0
0.1 cN
40
PA: 0.12 mm Ø
DTVB-500
DTVX-500
5.0 - 500.0
0.1 cN
40
PA: 0.20 mm Ø
DTVB-1000
DTVX-1000
50 - 1000
1 cN
40
PA: 0.30 mm Ø
DTVB-2000
DTVX-2000
200 - 2000
1 cN
40
PA: 0.50 mm Ø
Dimensions of the measuring head
Guide Rollers
Order-Code
Line Speed
...m/min
Roller Material
Standard
900
Hardcoated aluminium
Code K
2000
Hardcoated aluminium
Code T
450
Plastic (POM) black
Code W
450
Nickel-plated steel
Specifications
Model DTVB:
Calibration:
According to SCHMIDT factory procedure
Accuracy:
From 10 % up to 90 % of measuring range: ±0.5 % full scale and ±1 digit remaining tension range and other calibration material: ±3 % full scale and ±1 digit
Overranging:
Approx. 15 % full scale, without accuracy guarantee
Overload protection:
100 % of tension range
Measuring principle:
Strain gauge bridge
Measuring roller deflection:
0.2 mm max.
Signal processing:
Digital
Measuring frequency:
62 measurements/s
Converter:
12 bit A/D
Temperature coefficient:
Gain: less than ±0.01 % full scale/ºC Zero point: less than ±0.3 % full scale/ºC
Display:
LCD 4 digit, 12 mm high
Display update rate:
Adjustable 0.5, 1, 2 oder 4 s (averaging)
Field adjustment:
±7 steps in 1.5 % increments
Memory:
MIN, MAX, PEAK and last value
Temperature range:
10 - 45 ºC
Air humidity:
85 % RH, max.
Power supply:
4 batteries 1.5 V, size AA (approx. 20 h continuous use)
Auto power off:
After 2 min of non-use
Housing material:
Die-cast aluminium
Housing dimensions:
235 x 76 x 45mm (LxWxH)
Weight, net gross):
Approx. 680 g (approx. 1500 g)
Expanded memory:
Additional specifications for model DTVX:
Max. 100 readings, MAX, MIN, and PEAK values as well as average and standard deviation
Sigital output signal:
RS 232 C (4800, 8, N, 2)
Analog output signal:
0-1 V DC (converter rate 16 ms)
Digimatic interface:
Mitutoyo
Notation
ABTEK CORP.
7F., No.42, Sec. 1, Chang -An E. Rd., Zhongshan Dist., Taipei City 104, Taiwan (R.O.C.)